| Co2 Accuracy | 0 – 40 mmHg, ±2 mmHg |
| 41 – 70 mmHg, ±5% of reading | |
| 71 – 100 mmHg, ±8% of reading | |
| 101 – 150 mmHg, ±10% of reading | |
| NOTE: Accuracy is based upon the following conditions: Fully hydrated standard gas mixture and a gas Temperature of 35° C. | |
| Co2 Measurement Range | 0 to 150 mmHg, 0 to 19.7%, 0 to 20 kPa |
| Co2 Noise | RMS noise of the sensor is ≤ to 0.25 mmHg at 7.5% CO2 |
| Co2 Resolution | 0.10 mmHg – 0 to 69 mmHg |
| 0.25 mmHg – 70 to 150 mmHg | |
| Co2 Stability | Short Term Drift: Drift over four hours shall not exceed 0.8 mmHg maximum, measured with 5% CO2. |
| ong Term Drift: Accuracy specification will be maintained over a 120 hour period, measured with 5% CO2. | |
| Sampling Rate | 100 Hz |
| Respiration Accuracy | ±1 breath |
| Respiration Range | 2 to 150 breaths per minute (BPM) |